Thermal shock chamber


Introduction of ventilation-type (three rooms) thermal shock chamber: this chamber is suitable for the product (complete machine), parts, materials to estimate the ability of bearing speedy temperature change. This test can know the influence caused by the temperature variation one or more times on the test samples. The main parameters to influence the temperature change test have the max value and min value of temperature change range, the hold time of the samples under high temperature and low temperature, the number of test cycle, etc.
 
CJS3 series technical parameters
Type CJ602S3 CJ603S3 CJ605S3
 
Performance parameters
Preheating
room
Upper limit of preheating temperature 200℃
Temperature increasing time R.T.→+200℃约40min
Pre-cooling room Lower limit of pre-cooling temperature 0~-75℃
Temperature decreasing time R.T.→-70℃约90min
Workroom Thermal shock range High temperature 60℃~+150℃;low temperature -55℃~-10℃
Temperature fluctuation ≤±0.5℃
Temperature deviation ≤±2℃
Temperature recovering time ≤5min
 
Temperature recovery conditions
CJ602S3I condition 1 exposure in high temperature 125℃,exposure in low temperature -40℃,exposure duration 30min,sample weight 10Kg Al
condition 2 30min exposure in high temperature 150℃, 5min in environment temperature, 30min exposure in low temperature -55℃,sample weight 10Kg Al
CJ602SII condition 1 exposure in high temperature 125℃,exposure in low temperature -40℃,exposure duration 30min, sample weight 20Kg Al
condition 2 30min exposure in high temperature 150℃, 5 min in environment temperature,30min exposure in low temperature -55℃,sample weight 20Kg Al
condition 3 exposure in high temperature +150℃,exposure in low temperature -55℃,exposure duration 30min, sample weight 10Kg Al
CJ603S3I condition 1 exposure in high temperature 125℃,exposure in low temperature -40℃,exposure duration 30min, sample weight 20Kg Al
condition 2 30min exposure in high temperature 150℃,5 min in environment temperature,30min exposure in low temperature -55℃,sample weight 10Kg Al
CJ603S3II condition 1 exposure in high temperature 125℃,exposure in low temperature -40℃,exposure duration 30min, sample weight 20Kg Al
condition 2 30min exposure in high temperature 150℃, 5 min in environment temperature, 30min exposure in low temperature -55℃, sample weight 20Kg Al
condition 3 exposure in high temperature 150℃,exposure in low temperature -55℃,exposure duration 30min, sample weight 10Kg Al
CJ605S3I condition 1 exposure in high temperature 125℃,exposure in low temperature -40℃,exposure duration 30min, sample weight 10Kg Al
condition 2 30min exposure in high temperature 150℃, 5 min in environment temperature, 30min exposure in low temperature -55℃, sample weight 10Kg Al
CJ605S3II condition 1 exposure in high temperature 125℃,exposure in low temperature -40℃,exposure duration 30min, sample weight 20Kg Al
condition 2 30min exposure in high temperature 125℃, 5 min in environment temperature, 30min exposure in low temperature -55℃,sample weight 20Kg Al
condition 3 30min exposure in high temperature 150℃, 30min exposure in low temperature - 55℃,load 10Kg Al
 
Structure
Shell Cold-rolled steel plate with surface spray(ivory)
Inner wall material Stainless steel plate
 
Thermal insulation material Polyurethane foam+ superfine glass wool
 
Refrigeration Refrigerating method
 
Water cooling
Refrigerator
 
Imported compressor
Test hole one Φ50mm at the left side
Controller Siemens PLC controller and Siemens color LCD touch screen.
Data storage and functional interface Native SD data card, RS485 interface, USB interface
Workroom dimension (mm) D 670 670 670
W 650 970 970
H 460 460 780
 
External dimension (mm)
D 2245 2245 2245
W* 1610 1940 1940
H 2050 2050 2470
Installed power(KW) Type I 35KVA 50KVA 50KVA
Type II
 
38KVA 53KVA 53KVA
Power supply AC380V 50Hz three phase four wire + ground wire and grounding protection resistance is less than 4Ω
Standard configuration 1 product operation manuals. 1 test report, 1 certification, 1 quality certificate, 5 printing papers, 2 shelves and 4 racks, 1 ribbon and 1  silicone rubber soft plug with castors
Satisfy the standards GJB 150.3、GJB 150.4、GJB150.5、GB/T 2423.1、GB/T 2423.2
 
 
 
Chongqing Yinhe Experimental Equipment CO.,LTD. 2012 渝ICP备09054543号 技术支持:Hansa